		<id>https://bionmr.unl.edu/mediawiki/mediawiki/index.php?title=1H_NMR_Analysis&amp;diff=238&amp;oldid=prev</id>
		<title>Bozhang: Created page with &quot;===ACDLab/SpecManager===  ==Import spectral data==  Use 1D Macro to load the desired series of 1D 1H NMR &quot;.fid&quot; data files. When this is done, by default, the “group treatme...&quot;</title>
		<link rel="alternate" type="text/html" href="https://bionmr.unl.edu/mediawiki/mediawiki/index.php?title=1H_NMR_Analysis&amp;diff=238&amp;oldid=prev"/>
		<updated>2012-09-25T05:36:46Z</updated>

		<summary type="html">&lt;p&gt;Created page with &amp;quot;===ACDLab/SpecManager===  ==Import spectral data==  Use 1D Macro to load the desired series of 1D 1H NMR &amp;quot;.fid&amp;quot; data files. When this is done, by default, the “group treatme...&amp;quot;&lt;/p&gt;
&lt;p&gt;&lt;b&gt;New page&lt;/b&gt;&lt;/p&gt;&lt;div&gt;===ACDLab/SpecManager===&lt;br /&gt;
&lt;br /&gt;
==Import spectral data==&lt;br /&gt;
&lt;br /&gt;
Use 1D Macro to load the desired series of 1D 1H NMR &amp;quot;.fid&amp;quot; data files. When this is done, by default, the “group treatment” is on and ACD operations will be applied to the entire group. &lt;br /&gt;
&lt;br /&gt;
Steps:&lt;br /&gt;
1. Click on the “1D NMR Macro” button.&lt;br /&gt;
&lt;br /&gt;
2. On the “Execute Group Window”, make sure:&lt;br /&gt;
&lt;br /&gt;
a) “Plate” should not be checked.&lt;br /&gt;
&lt;br /&gt;
b) Check “spectra”.&lt;br /&gt;
&lt;br /&gt;
c) Click on “add group”. The wild cards (&amp;quot;*&amp;quot; for numbers, &amp;quot;?&amp;quot; for letters) can be inserted in the directory mask line. The file mask should read “.fid”. Click “Ok”.&lt;br /&gt;
&lt;br /&gt;
3. When all of the data files are correctly listed, click “Ok” to execute the group macro and load the files into ACDLab/SpecManager. Assuming that the files loaded correctly, click “Ok” when the process if finished.&lt;br /&gt;
&lt;br /&gt;
==Pre-process the spectra==&lt;br /&gt;
&lt;br /&gt;
If appropriate, apply a window function. To match what TOPSPIN does automatically when processing data on the 500 MHz console, use a 0.3 Hz exponential window function prior to Fourier transform (FT).  &lt;br /&gt;
&lt;br /&gt;
Steps:&lt;br /&gt;
&lt;br /&gt;
1. Click on “WFunctions” in the operations bar.  &lt;br /&gt;
&lt;br /&gt;
2. Set LB = 0.3 Hz&lt;br /&gt;
&lt;br /&gt;
3. Choose “Exponential”.&lt;br /&gt;
&lt;br /&gt;
4. Click “Ok”.&lt;br /&gt;
&lt;br /&gt;
5. Fourier transform the FID files&lt;br /&gt;
&lt;br /&gt;
6. Click on “Fourier Tr.” in the operations bar.  &lt;br /&gt;
&lt;br /&gt;
7. Phase correct the spectra.&lt;br /&gt;
&lt;br /&gt;
8. Click on the “Phase Corr.” button on the operations bar.&lt;br /&gt;
&lt;br /&gt;
9. Use the “Auto Simple” option.  &lt;br /&gt;
&lt;br /&gt;
10. Click on the “checkmark” to accept the result.  &lt;br /&gt;
&lt;br /&gt;
11. Set chemical shift reference using the TMSP peak.&lt;br /&gt;
&lt;br /&gt;
12. Click on the “Reference” button on the operations bar.&lt;br /&gt;
&lt;br /&gt;
13. Check the “Options” to verify that TMS is set.  Both TMSP and TMS have chemical shift values of 0.000 ppm.&lt;br /&gt;
&lt;br /&gt;
14. Click on “Auto”.  Then click “Ok” in response to the “There are no labeled peaks available…” message if it appears.  &lt;br /&gt;
&lt;br /&gt;
&lt;br /&gt;
==Integrate the spectra and export the peak list==&lt;br /&gt;
&lt;br /&gt;
Using the “auto integrate” option:&lt;br /&gt;
&lt;br /&gt;
&lt;br /&gt;
Steps:&lt;br /&gt;
&lt;br /&gt;
1. Click on the Integration operation.&lt;br /&gt;
&lt;br /&gt;
2. Check the “Options”&lt;br /&gt;
&lt;br /&gt;
a) Set for “Bucket Integration”&lt;br /&gt;
&lt;br /&gt;
b) Set width of buckets to 0.025 ppm&lt;br /&gt;
&lt;br /&gt;
c) Choose intelligent bucketing with width looseness (%) = 50 and method = sum&lt;br /&gt;
&lt;br /&gt;
d) Set reference to “whole spectrum”.  &lt;br /&gt;
&lt;br /&gt;
e) Click “Ok”&lt;br /&gt;
&lt;br /&gt;
f) Click on “Auto” to perform the integration&lt;br /&gt;
&lt;br /&gt;
g) Click on the “checkmark” to accept the result.&lt;br /&gt;
&lt;br /&gt;
&lt;br /&gt;
3. Use the Series pull-down menu to bring up the table of common integrals.  A right-click will allow one to save the integral table.  The saved &lt;br /&gt;
table can be imported directly into SIMCA or opened in Excel.&lt;/div&gt;</summary>